Other, other text: additional information: d10, h6, h10: device evaluation: one smiths medical medfusion pump was returned for analysis with the top and bottom cases in an excellent condition and no visible contamination.Event log history was performed and indicated that primary audible alarm post was recorded in history.During analysis, the reported issue was able to be duplicated.It was noted that the speaker was the cause of the reported issue.Service replaced the speaker to correct the reported issue.Service also performed preventive maintenance and all functional testing.Based on the evidence, the complaint was confirmed, and the problem source of the reported event is unknown.
|