Related manufacturer reference number: 2017865-2020-14739.It was reported the asymptomatic patient presented for an unrelated procedure.Upon device check, it was observed the right ventricular lead had low lead impedance and oversensing episodes with pacing inhibition.The implantable cardioverter defibrillator had over current detection alerts, oversensing with pacing inhibition, and shorted output stage detection episodes.The lead was capped and replaced on (b)(6) 2020 and a new device implanted.The patient was stable.
|
The analysis indicated the cause of the damage was high voltage delivery into a low impedance load.This low impedance load is consistent with a lead-to-can arc and, therefore, is not considered a device malfunction.Interrogation of the device revealed that the device was at elective replacement indicator and in bvvi (backup vvi) when received.Review of the device image showed that the device entered backup vvi mode on sep 14, 2020 due to a por (power on reset).The cause of the por (power on reset) is also consistent with a lead-to-can arc and, therefore, is also not considered a device malfunction.All low voltage device functions were normal.
|