Model Number 9600TFX26
Device Problem Failure to Align (2522)
Patient Problem Injury (2348)
Event Date 09/09/2020
Event Type  Injury  
Event Description
As reported from our affiliates in belarus, during a transfemoral transcatheter aortic valve replacement (tavr) with a 26mm sapien 3 valve, the complex vessels anatomy of the patient (tortuous and kinked vessels) led to a non-coaxial alignment of the delivery system with the valve in the aortic annulus.During valve inflation, the valve moved into the aorta.   the patient was converted to a conventional surgery.The patient was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key10624019
MDR Text Key209764342
Report Number2015691-2020-13834
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeBO
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/21/2021
Device Model Number9600TFX26
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/09/2020
Initial Date FDA Received10/02/2020
Date Device Manufactured10/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age82 YR
Patient Weight75
-
-