Model Number 71360020
Device Problem Mechanical Jam (2983)
Patient Problem No Information (3190)
Event Date 09/08/2020
Event Type  Injury  
Event Description
It was reported that the accord tensioner cable got a wire caught in surgery.A change in surgical technique was reported.It is unknown how the procedure was completed or if there was any patient harm.A delay of 30 minutes or less was reported.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirms there is a wire seized inside the accord tensioner and can¿t be removed.This device was manufactured in 2010.This device shows signs of significant wear/use.A medical investigation will be performed.Proceed based on information provided/available for the investigation; if no relevant clinical information is provided, recommend closure.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORD TENSIONER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10624720
MDR Text Key209744285
Report Number1020279-2020-05147
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010485359
UDI-Public03596010485359
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71360020
Device Catalogue Number71360020
Device Lot Number10JMC7030
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Initial Date Manufacturer Received 09/08/2020
Initial Date FDA Received10/04/2020
Supplement Dates Manufacturer Received10/16/2020
Supplement Dates FDA Received10/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-