Based on our investigation, we can conclude that the trajectory of the guide aligns with the final plan.Additionally, all production processes were found to have been properly followed.The trajectory deviation may have been caused by an inaccurate patient scan, and/or drill slippage occurring due to the change in morphology caused by a tooth extraction.
|
The doctor used the guide for implant surgery and placed implants at sites #3, #4, and #13.Three weeks after surgery, the patient went back to the doctor's office with some swelling around site #13.The doctor prescribed antibiotics, but after the fourth week of swelling, the doctor decided to take a new ct scan.From the scan, the doctor observed that #13 had perforated the buccal plate.The implant was then removed and the site was grafted.No issues were reported with implants #3 and #4.
|