• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 15CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 15CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66000375
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/08/2020
Event Type  malfunction  
Event Description
It was reported that the adhesive strength is too low.It is unknown if there was a patient involved.
 
Manufacturer Narrative
We have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition, it can be confirmed that all finished product specification testing was satisfied at the point of release.A complaints history review was carried out using the lot and part numbers provided, there have been further complaints reported with this failure mode in the past three years.The device was used for treatment.It was reported that the dressing did not adhere to the patient¿s skin satisfactorily.As no samples were returned, a thorough product evaluation could not be carried out.As with all adhesive products, the skin site must be thoroughly cleaned and dried prior to application.If there is any moisture on the skin surrounding the catheter, the adhesive performance of this dressing may be compromised.If the dressing becomes wet whilst applied, it may have to be removed and a new dressing applied.We have not been able to confirm a relationship between the event and the device or determine a root cause on this occasion.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE FLEXIFIX 15CMX10M CTN 1
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key10632065
MDR Text Key209942907
Report Number8043484-2020-03131
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number66000375
Device Lot Number2004
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/11/2020
Initial Date FDA Received10/05/2020
Supplement Dates Manufacturer Received10/27/2020
Supplement Dates FDA Received11/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-