• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. REVEAL LINQ MOBILE MANAGER APP; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. REVEAL LINQ MOBILE MANAGER APP; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LMM-GENERAL
Device Problems Device Difficult to Program or Calibrate (1496); Computer Operating System Problem (2898)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/21/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the diagnostic mobile programmer application froze during programming process.The programming was unable to be completed due to the session freezing and searching for a bluetooth connection.Following multiple attempts to exit the session and re-do the insert session, the user was advised on troubleshooting steps which involved using ¿¿check device¿¿.The user was then able to view the device, but the patient demographics were gone.The user then turned on the required parameters and ended the session.The device was then checked again to ensure everything was on and functioning.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ MOBILE MANAGER APP
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10634520
MDR Text Key209998957
Report Number2182208-2020-02240
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLMM-GENERAL
Device Catalogue NumberLMM-GENERAL
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/21/2020
Initial Date FDA Received10/06/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age80 YR
-
-