Model Number TSH
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/01/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The customer's calibration and qc data were requested but not provided.The observed differences in tsh values generated with the roche assay and abbott architect assay are most likely caused by differences in the overall setups of the assays, the antibodies used, differences in reference materials, and the differences in the standardization methodology used.Based on the available data, a general reagent issue could be excluded.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Event Description
The initial reporter received questionable elecsys tsh assay results for one patient tested on a cobas 6000 e 601 module serial number (b)(4).The questionable results were reported outside the laboratory, but the physician asked for re-measurement of the patient¿s sample.A sample from the patient was submitted for investigation and was tested on a cobas 8000 e 602 module and an abbott architect analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH ASSAY
Type of Device
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10634795
MDR Text Key209998131
Report Number1823260-2020-02449
Device Sequence Number1
Product Code JLW
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K961491
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTSH
Device Catalogue Number11731459122
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/15/2020
Initial Date FDA Received10/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age57 YR
-
-