The customer contacted a siemens customer care center (ccc).Quality controls failed, after which the customer noticed they did not remove the rinse float from the barbicide container after performing tube disinfection on their bcs xp system.The customer set the rinse float back in the rinse bottle, replaced the reagents, and performed rinses.Quality controls were then rerun and recovered within acceptable ranges.The instrument is now fully operational.The cause of the event is use error.The system is performing according to specifications.No further evaluation of this device is required.
|
A discordant, falsely low d-dimer result was obtained on a bcs xp system using innovance d-dimer reagent.The discordant result was reported to the physician(s).The same sample was repeated for d-dimer two times on an alternate bcs xp system, once using the d-dimer low application and once using the d-dimer high application.Both results recovered higher than the initial discordant result.The patient was redrawn and the redrawn sample was run for d-dimer on the same alternate bcs xp system, once using the d-dimer low application and once using the d-dimer high application.Both results recovered higher than the initial discordant result.The higher repeat results were reported, as the correct results, to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low d-dimer result.
|