Device evaluated by manufacturer: the device was returned for analysis.The handshake connection, sheath, burr, and coil were microscopically and visually examined.Inspection of the device revealed that the sheath was separated 22.3cm distal of the strain relief with blood in the sheath.Majority of the coil was stretched, and the handshake connection was received stuck in the housing.The housing was dismantled to check for damage inside the housing and no damage was present.The damage to the sheath is consistent to damage from over tightening the hemostasis valve.
|