Model Number 0673 |
Device Problems
Signal Artifact/Noise (1036); High impedance (1291); Failure to Sense (1559)
|
Patient Problem
No Code Available (3191)
|
Event Date 09/02/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4) captures the reportable event stating that the patient had a prolonged procedure.At this time, the product has not been returned.If the product is returned, analysis will be performed and would be updated at that time should pertinent information be provided.
|
|
Event Description
|
It was reported that during the implant procedure, when this right ventricular (rv) lead was placed, only a flat or noisy line was seen, and no signals were retrieved from the lead.The physician tried with several new cables and a different pacing system analyzer (psa), however, this didn't resolve the issue.When the pacing mode was enabled, the pacing impedance measurements were greater than 4000 ohms.The physician decided to remove this lead and replace it.No adverse patient effects were reported.
|
|
Event Description
|
It was reported that during the implant procedure, when this right ventricular (rv) lead was placed, only a flat or noisy line was seen, and no signals were retrieved from the lead.The physician tried with several new cables and a different pacing system analyzer (psa), however, this didn't resolve the issue.When the pacing mode was enabled, the pacing impedance measurements were greater than 4000 ohms.The physician decided to remove this lead and replace it.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Patient code 3191 captures the reportable event stating that the patient had a prolonged procedure.Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Testing was completed to assess lead electrical performance and inner/outer insulation integrity.Measurements throughout these tests were within normal limits.Laboratory testing was unable to reproduce the reported clinical observations.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported high impedances, noise, and failure to sense.
|
|
Search Alerts/Recalls
|
|