• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS P 512 PRE-ANALYTICAL SYSTEM; AUTOMATED PREANALYTICAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS P 512 PRE-ANALYTICAL SYSTEM; AUTOMATED PREANALYTICAL SYSTEM Back to Search Results
Model Number P512
Device Problem Failure to Align (2522)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/09/2020
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).The field service engineer discovered a worn and damaged belt.He replaced the belt, and performed system adjustments.The investigation determined the service actions resolved the issue.
 
Event Description
The initial reporter complained of sample contamination during the sorting of sample tubes on a cobas p 512 pre-analytical system.The reporter stated during the sorting process, an open sample tube hit a rack, which caused serum to be splattered, and contaminated a sample two positions ahead.The customer observed this event, and removed the contaminated sample from the rack, and the pre-analytical system.The customer confirmed no erroneous results, or serious injuries occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS P 512 PRE-ANALYTICAL SYSTEM
Type of Device
AUTOMATED PREANALYTICAL SYSTEM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key10641765
MDR Text Key210185999
Report Number1823260-2020-02469
Device Sequence Number1
Product Code JQP
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberP512
Device Catalogue Number05083435001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/10/2020
Initial Date FDA Received10/07/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-