Catalog Number UNK HIP FEMORAL STEM |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Edema (1820); Pain (1994); Test Result (2695); No Code Available (3191)
|
Event Date 03/20/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).Initial reporter occupation is lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Medical record received.After review of medical record, patient was revised to address failed left total hip arthroplasty.Revision notes stated that the hip joint there was synovial fluid which appears slightly tented of gray matter.Head and liner were removed.Patient experienced pain and elevated metal ion.Doi: (b)(6) 2006 - dor: (b)(6) 2019 (left hip).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.
|
|