• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00516730
Device Problem Migration (4003)
Patient Problem No Code Available (3191)
Event Date 09/17/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).The complainant indicated that the device remains implanted and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2020 that a wallflex esophageal fully covered stent was implanted to treat a 3cm malignant stricture in the esophagus during an esophagogastroduodenoscopy (egd) with stent placement performed on (b)(6) 2020.Reportedly, the patient's anatomy was not tortuous and was not dilated prior to stent placement.According to the complainant, two days after the stent placement during a routine follow-up procedure, a nasogastric tube was placed and it was noted under kidney, ureter, bladder (kub) imaging that the stent had migrated.On (b)(6) 2020, an egd procedure was performed and the stent was repositioned with rat tooth forceps and sutured in place.The stent remains implanted and the procedure was completed with original stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX ESOPHAGEAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key10652751
MDR Text Key210570765
Report Number3005099803-2020-04507
Device Sequence Number1
Product Code ESW
UDI-Device Identifier08714729778066
UDI-Public08714729778066
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K091510
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/02/2021
Device Model NumberM00516730
Device Catalogue Number1673
Device Lot Number0025557094
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/17/2020
Initial Date FDA Received10/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/03/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age70 YR
Patient Weight47
-
-