It was reported that while reaming the acetabulum, the reamer shaft broke at the connection to the reamer hand piece due to excessive force.The device broke during surgery, while inside the patient.Procedure was completed using a sn backup device.No delay or injury to the patient was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H3, h6: the device, used in treatment, was returned for evaluation.A visual inspection of the returned device confirms the u joint fractured on the inner portion of the device.All pieces were returned for evaluation.This device shows signs of significant wear/usage.A medical investigation was conducted and confirms the single photo provided confirm the reported failure.Based on the information provided the root cause of the device breakage was due to excessive force.The patient impact beyond the reported instrument breakage and possible retained foreign body of a non-implantable device could not be determined, although, corrosion/local irritation and possible migration of possible retained foreign body could not be ruled out.Per communications, the procedure was completed using a sn backup device.Since there was, no delay or injury to the patient, no further clinical/medical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|