• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 26 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. COCR 12/14 FEM HEAD 26 + 0; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 71302600
Device Problem Unintended Movement (3026)
Patient Problems Injury (2348); Joint Dislocation (2374)
Event Date 09/11/2020
Event Type  Injury  
Event Description
It was reported the first revision tka surgery was on (b)(6) 2018.Patient suffered dislocation of femoral head and retainer ring from acetabular shell and constrained liner construct.Primary surgery date unknown.Second revision (b)(6) 2020.No injury reported except need for revision.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that the root cause of the first episode of dislocations cannot be concluded.The length of time insitu with normal ware may have contributed to the issue.The stated poor gluteal musculature and the reported ¿traumatic event of a forward flexion and the popping of the device¿ may have contributed to the cause of the second episode dislocations.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file and instructions for use documents revealed this failure mode and potential harm was previously identified.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved or product information, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COCR 12/14 FEM HEAD 26 + 0
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10653064
MDR Text Key210583235
Report Number1020279-2020-05282
Device Sequence Number1
Product Code LPH
UDI-Device Identifier03596010194077
UDI-Public03596010194077
Combination Product (y/n)N
PMA/PMN Number
K963509
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71302600
Device Catalogue Number71302600
Device Lot Number17AM22744
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/11/2020
Initial Date FDA Received10/08/2020
Supplement Dates Manufacturer Received06/26/2021
Supplement Dates FDA Received06/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age86 YR
-
-