• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ANTHOLOGY HO TR NECK SZ 1-6; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ANTHOLOGY HO TR NECK SZ 1-6; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71365801
Device Problems Material Disintegration (1177); Naturally Worn (2988)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/14/2020
Event Type  malfunction  
Event Description
It was reported during the procedure, thr, that the coating on trial necks wear off, causing that the magnet detaches from the trial neck.The procedure was completed without delay, using the same device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
Updated d4: lot number and exp / mnf date.
 
Manufacturer Narrative
H10: h6: the associated device, used in treatment, were returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device's coating is worn off, causing magnet mechanism to malfunction.The device was manufactured in 2013 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY HO TR NECK SZ 1-6
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10655466
MDR Text Key210606258
Report Number1020279-2020-05289
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010559876
UDI-Public03596010559876
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71365801
Device Catalogue Number71365801
Device Lot Number13BM02880
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2020
Initial Date Manufacturer Received 09/14/2020
Initial Date FDA Received10/09/2020
Supplement Dates Manufacturer Received09/14/2020
01/11/2021
Supplement Dates FDA Received11/18/2020
01/12/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-