• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Catalog Number 425-00
Device Problem Complete Loss of Power (4015)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2020
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A verification of the reported failure mode was conducted and 8 devices were taken from current production (425-00 concha neptune lot #: 73j200005n) at the manufacturing facility and were functionally tested.No issues were encountered during the functional testing.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
The complaint is reported as: the unit shuts off during use.No patient injury or consequence reported.The condition of the patient is reported as "fine".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key10656461
MDR Text Key210645404
Report Number3003898360-2020-00806
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 09/21/2020
Initial Date FDA Received10/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-