• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HUDSON CONCHA NEPTUNE; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Catalog Number 425-00
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/22/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
The complaint is reported as: "service light/high temp".It is unknown if the issue was detected during use on a patient.Several attempts were made to get further clarification on the description; however, the customer has not responded.
 
Manufacturer Narrative
Qn#(b)(4).The sample was returned for evaluation.A visual exam was performed and there were no defects observed.Functional testing was performed and the unit was connected to 110vac.The unit passed the initial power connect test and navigated through the power-on self-test (p.O.S.T.) with no issues.During the temperature display accuracy test the unit displayed the correct temperatures and properly alarmed in the high temperature scenario.The unit was prepared for the functional bench test where water, an adult breathing circuit (880-36kit), 2-3 lpm of air pressure, and a 382-10 universal water column was connected to the unit for a real time operational scenario.The settings were set to full rainout, invasive, at 37 degrees c.The unit functioned without any interruption or functional anomalies for ~4.5 hours.Based on the investigation performed, the reported complaint could not be confirmed.Functional testing did not reveal any operational anomalies.
 
Event Description
The complaint is reported as: "service light/high temp".It is unknown if the issue was detected during use on a patient.Several attempts were made to get further clarification on the description; however, the customer has not responded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CONCHA NEPTUNE
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
MDR Report Key10656630
MDR Text Key210650284
Report Number3003898360-2020-00830
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K131912
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number425-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2020
Initial Date Manufacturer Received 09/22/2020
Initial Date FDA Received10/09/2020
Supplement Dates Manufacturer Received11/10/2020
Supplement Dates FDA Received11/10/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-