The patient was undergoing a thrombectomy procedure in the right internal jugular using the lightning aspiration tubing (lightning), indigo system cat12 aspiration catheter (cat12), and penumbra engine (engine).During the procedure, the physician advanced the cat12 to the face of the clot and turned the power on the engine; however, there was no aspiration.It was reported that the lights around the power button on the engine were illuminated but the engine was unable to produce aspiration.Therefore, the engine was removed.The procedure was completed using a penumbra system aspiration pump max 110 (pump max).There was no report of an adverse effect to the patient.
|
Please note that the following sections were incorrectly reported on the initial mfr report and are being corrected on this follow-up #01 mfr report: 3005168196-2020-01685 1.Section h.Box 5.Labeled for single use? results: the engine was able to power on, but no vacuum pressure was observed.Only the power indicator lights, and the cooling fans were turned on.Therefore, the device housing was opened and observed that the ribbon cable wires leading to the vacuum pump were damaged.Conclusions: evaluation of the returned engine confirmed that the device was unable to produce aspiration.During functional testing, the power indicator lights, and cooling fans were able to be activated but no vacuum was observed.Therefore, the device housing was opened, and it was observed that the wires leading to the vacuum pump were damaged.The damage to the wires likely contributed to the engine unable to produce aspiration during the procedure.Penumbra engines are inspected at incoming quality control which includes a visual inspection as well as a verification of test results to ensure specifications for each output are met.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|