• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THERAKOS, INC. CELLEX PHOTOPHERESIS SYSTEM Back to Search Results
Model Number NOT APPLICABLE
Device Problem Fluid/Blood Leak (1250)
Patient Problem Insufficient Information (4580)
Event Date 09/11/2020
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pump tubing organizer (pto) leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot j212 was conducted.There were no non-conformance's associated with this lot.This lot met all release requirements.A review of kit lot j212 shows no trends.Trends were reviewed for complaint category, pto leak.No trends were detected for this complaint category.Photographs were provided by the customer for evaluation.The complaint kit with smart card was not returned.A review of the photographs verify a blood leak coming from the pump tubing organizer (pto) installed on the cellex instrument pump deck.The leak appears to be coming from inside the pto; however, the exact location of the leak could not be determined based on the photographs provided.A device history record (dhr) review did not result in any related non-conformance's and this kit lot passed all lot release testing.The root cause for the pto leak could not be determined based on the available information.No further action is required at this time.This investigation is now complete.(b)(4).P.T.(b)(6) 2020.
 
Event Description
The customer contacted mallinckrodt to report they experienced a pump tubing organizer (pto) leak with their cellex photopheresis kit ("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported a blood leak was observed in the pump tubing organizer during the buffy coat collection phase of the procedure.The customer aborted the ecp treatment and did not return residual blood within the kit to the patient.The patient was reported to be in good condition.The customer discarded the kit and returned photographs for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin, D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key10658330
MDR Text Key212776626
Report Number2523595-2020-00113
Device Sequence Number1
Product Code LNR
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2022
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXECP
Device Lot NumberJ212
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/11/2020
Initial Date FDA Received10/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
Patient Weight63
-
-