• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPACELABS HEALTHCARE INC. SPACELABS ULTRAVIEW SL CENTRAL MONITOR; UVSL CENTRAL MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPACELABS HEALTHCARE INC. SPACELABS ULTRAVIEW SL CENTRAL MONITOR; UVSL CENTRAL MONITOR Back to Search Results
Model Number 91387-38
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Tech support confirmed the hospital network went down.Onsite investigation by hospital it, performed network router reboot to restore the reported issue.Users confirm they have not had any issues.This report is considered final and the issue closed.
 
Event Description
Spacelabs received a report that on (b)(6) 2020, lost visibility of hardwire and telemetry patients locally and remotely.No injury was reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACELABS ULTRAVIEW SL CENTRAL MONITOR
Type of Device
UVSL CENTRAL MONITOR
Manufacturer (Section D)
SPACELABS HEALTHCARE INC.
35301 se center st.
snoqualmie WA 98065
Manufacturer Contact
marc rivas
35301 se center st.
snoqualmie, WA 98065
4253635568
MDR Report Key10659451
MDR Text Key224670209
Report Number3010157426-2020-00039
Device Sequence Number1
Product Code MHX
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K102422
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/01/2005,10/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number91387-38
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 01/14/2005
Initial Date FDA Received10/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-