• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA PK 4.5MM W/2 UB -WHT &BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA PK 4.5MM W/2 UB -WHT &BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202595
Device Problems Break (1069); Material Separation (1562)
Patient Problems No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/21/2020
Event Type  Injury  
Event Description
It was reported that during a rotator cuff repair, the anchor broke while screw-in.The broken pieces were removed with tweezers.The procedure was successfully completed without significant delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Event Description
It was reported that during a rotator cuff repair, the anchor broke while screw-in.The broken pieces were removed with tweezers.The procedure was successfully completed without significant delay using a back-up device in the originally drilled bone hole.No other complications were reported.Patient's current status is good.
 
Manufacturer Narrative
H10: internal complaint reference: (b)(4).H2: additional information ¿h6: health effect - clinical code and impact code, medical device problem code.¿ h3, h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer, found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A visual inspection of the returned device found that it is not in its original packaging.The insertion device, suture material and a fractured anchor were returned.All components have debris on them.Based on the condition of the product material found during visual inspection, additional material testing is not required.Per case details, the broken anchor was retrieved from the patient.The procedure was completed using a back-up device.No patient injuries or adverse consequences were reported.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA PK 4.5MM W/2 UB -WHT &BL
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10667255
MDR Text Key211247750
Report Number1219602-2020-01565
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010647467
UDI-Public03596010647467
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K093228
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/01/2024
Device Model Number72202595
Device Catalogue Number72202595
Device Lot Number2034465
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2020
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/22/2020
Initial Date FDA Received10/12/2020
Supplement Dates Manufacturer Received05/04/2022
11/02/2022
Supplement Dates FDA Received05/06/2022
11/03/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age25 YR
Patient SexMale
Patient Weight69 KG
-
-