U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR LINERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN HIP ACETABULAR LINERS Back to Search Results
Catalog Number UNK HIP ACETABULAR LINER
Device Problem Naturally Worn (2988)
Patient Problems Calcium Deposits/Calcification (1758); Injury (2348); Osteolysis (2377); Test Result (2695); No Code Available (3191)
Event Date 08/20/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint #: (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Medical record received.After review of medical record, patient was revised to address synovitis with osteolysis.Revision notes stated that metal staining was noted.The source of metal staining appeared to be small amount of impingement in the posterosuperior neck of femoral component would impinge only with maximum external rotation and slight abduction.There was no impingement with regular extension, external rotation.Small divot on the femoral neck of the stem was noted and did not appear to be nearly big enough to warrant removing the well-fixed femoral component.If the liner had been poly instead of metal, certainly no metallic notching would have occurred.The posterior capsule was intact, except for a small hole about 1 cm in diameter.Extensive bone loss was noted.Manhole cover and liner were removed.After thorough debris irrigation, a manhole cover was placed and a new polyethylene liner was impacted.Heterotopic bone was debrided.Doi: (b)(6) 2006; dor: (b)(6) 2019; (left hip).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP ACETABULAR LINERS
Type of Device
HIP ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10670711
MDR Text Key211237538
Report Number1818910-2020-22169
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP ACETABULAR LINER
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/25/2020
Initial Date FDA Received10/13/2020
Supplement Dates Manufacturer Received10/19/2020
Supplement Dates FDA Received10/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age68 YR
-
-