• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH SOMATOM FORCE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH SOMATOM FORCE; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 10742326
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter's telephone number is (b)(6).Siemens completed a technical investigation of the reported event.The investigation concluded that no device malfunction occurred since the spiral was suspended by the user before the minimal data needed for the image reconstruction was acquired.The root cause is use error.No further action is warranted at this time.
 
Event Description
It was reported to siemens that a four year-old male patient had to be rescanned with the somatom force system because the images from the first scan could not be reconstructed.The user stated the scan of the patient's inner ear using a very short spiral to avoid dose for the lenses by suspending the scan range earlier.There was no injury to the patient aside from an additional x-ray dose.The health status of the patient is stable and was not adversely affected by the missing images.This report has been submitted with an abundance of caution.The reported event occurred in (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOMATOM FORCE
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forchheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forchheim, 91301
GM   91301
Manufacturer Contact
rebecca tudor
40 liberty blvd.
65-1a
malvern, PA 19355
6104486484
MDR Report Key10672953
MDR Text Key220740719
Report Number3004977335-2020-47797
Device Sequence Number1
Product Code JAK
UDI-Device Identifier04056869006956
UDI-Public04056869006956
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K190578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10742326
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2020
Initial Date FDA Received10/13/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age4 YR
-
-