Model Number 5201260 |
Device Problems
Electrical /Electronic Property Problem (1198); Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/24/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned to quest medical for evaluation.A follow-up medwatch will be submitted if additional information becomes available.
|
|
Event Description
|
A report was received regarding an alleged issue encountered with the console.The report states that the console displayed ec179 during bypass.The console was swapped out and the case was resumed.There were no patient complications.
|
|
Manufacturer Narrative
|
The console was evaluated and the resistor in the motor driver board was identified as the cause of the reported issue.The motor driver board was replaced and the console passed all operational verification tests.
|
|