Device evaluated by mfr: the returned product consisted of an ffr comet pressure wire connected to the occ cable.The tip, device shaft and sensor port were examined for damage or any irregularities.The wire shaft showed 2 kinks located 38cm and 139cm from the tip.Peeled coating on the shaft was noticed at the 139cm location.The tip showed bend damage.The occ cable was connected to the ffr link for signal verification.The signal was not present as designed.The wire was gently shaken to see if the sensor would move within the sensor housing and the sensor did move which verifies the sensor was detached from the fiber optic.The coefficient values were confirmed to be programmed.The wire was removed from the occ handle with no issues.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
It was reported that signal issues with a pressure guidewire occurred.During preparation and while outside the patient, a comet device was connected to the console, but no pressure signal occurred.It was noted that the guidewire did not light up at the console, and the light on the console was red.Another comet was used to complete the procedure.No patient complications were reported in relation to tis event.However, the device was returned and analysis revealed peeled coating.
|