Model Number 0600-0767 |
Device Problems
Erratic or Intermittent Display (1182); No Display/Image (1183); Electrical Shorting (2926)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device return is anticipated, however; at the time of the report the devices have not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
|
|
Event Description
|
The customer reported that during a patient procedure, using a glidescope core quickconnect cable, the cable was shorting out and they would lose the image.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
|
|
Manufacturer Narrative
|
D9, g3, g6, h2, h3, h6, h10 the glidescope core quickconnect cable was returned to verathon for evaluation.A verathon technical service representative evaluated the returned core quickconnect cable and could not reproduce the "intermittent image" issue; the unit functioned as intended.The glidescope core quickconnect cable was scrapped and a replacement was sent to the customer.No further investigation is required at this time.Verathon will continue to monitor for trends.
|
|