Model Number 866389
Device Problem Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused, or contributed to the event.A field service engineer was scheduled for the customer site, the customer cancelled the call.
 
Event Description
The customer reported that all the hosts in hospital are in local mode.The device was in use on a patient.There was no report of patient, or user harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE INFORMATION CENTER IX
Type of Device
CENTRAL STATION MONITOR
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
robert corning
3000 minuteman road
andover, MA 01810
9786871501
MDR Report Key10678221
MDR Text Key211345330
Report Number1218950-2020-06090
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102495
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/20/2020
Initial Date FDA Received10/14/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/13/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse