• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PROVISIONAL ARTICULAR SURFACE SIZE STRIPED YELLOW/E,F 10 MM HEIGHT; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PROVISIONAL ARTICULAR SURFACE SIZE STRIPED YELLOW/E,F 10 MM HEIGHT; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/15/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the provisional broke during trialing.No further information is available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay corrected and additional information.Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows that the trial provisional was fractured and all pieces were returned.The device also exhibits signs of usage.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause is considered to be determined as normal wear and tear from repeated use during the instrument's field life.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROVISIONAL ARTICULAR SURFACE SIZE STRIPED YELLOW/E,F 10 MM HEIGHT
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10682898
MDR Text Key211453041
Report Number0001822565-2020-03546
Device Sequence Number1
Product Code HWT
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00599503212
Device Lot Number60663005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/22/2020
Initial Date FDA Received10/15/2020
Supplement Dates Manufacturer Received10/16/2020
01/27/2021
Supplement Dates FDA Received11/10/2020
01/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-