ENCORE MEDICAL L.P. CLP HIP; HIP, STEM, CLP, LATERAL OFFSET, SIZE#5, IMPLANT, POLISHED NECK
|
Back to Search Results |
|
Model Number 426-11-050 |
Device Problems
Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 09/17/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
|
|
Event Description
|
Second revision surgery - implant (femoral stem) was loose, surgeon explanted and put in zimmer revision stem.
|
|
Manufacturer Narrative
|
The reason for this revision surgery was reported as due to loosening.The previous surgery and the surgery detailed in this event occurred 2 years and 10 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during the evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, patient bone deterioration, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implants safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
|
|
Search Alerts/Recalls
|
|
|