Model Number 5MAXJET7KIT-B
Device Problems Fluid/Blood Leak (1250); Fracture (1260)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2020
Event Type  malfunction  
Manufacturer Narrative
Results: the jet7 was kinked approximately 1.5 cm from the hub.The strain relief was unraveled, and a server kink was observed.Conclusions: evaluation of the returned jet7 revealed a server kink underneath the strain relief.This kink is likely the reported fracture.If the jet7 is forcefully mishandled at an extreme angle during use, damage such a kink may occur.During decontamination, the jet7 was flushed with air inside a syringe, and bubbles were observed coming from the kinked location underneath the strain relief.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure in the m1 segment of the middle cerebral artery (mca) using a penumbra system jet 7 reperfusion catheter (jet7), neuron max 6f 088 long sheath (neuron max), and non-penumbra microcatheter.During the procedure, while advancing the jet7 through the neuron max, the hospital technician noticed blood was leaking out from the proximal end near the hub of the jet7 and, therefore, it was removed.It was also reported that the jet7 may have been inadvertently fractured during preparation.The procedure was completed using a new jet7 and the same neuron max.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM JET7 REPERFUSION CATHETER
Type of Device
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key10684565
MDR Text Key211513697
Report Number3005168196-2020-01738
Device Sequence Number1
Product Code NRY
UDI-Device Identifier00815948020962
UDI-Public00815948020962
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K173761
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/19/2022
Device Model Number5MAXJET7KIT-B
Device Catalogue Number5MAXJET7KIT
Device Lot NumberF95098
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 09/24/2020
Initial Date FDA Received10/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
-
-