Internal complaint reference (b)(4).The reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.The device shows some wear from use.There is a small area of blue debris about halfway down the shaft.There is some deformation of the outer shaft at the distal end.A functional assessment of the device found the device opens and closes with difficulty and will not cut cleanly through test suture material.A complaint history review found no similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include maintenance issues, excessive force on the device, attempted correction of a damaged device, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|