• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TAEWOONG MEDICAL CO.,LTD. NITI-S ESOPHAGEAL COVERED STENT; ESOPHAGEAL STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TAEWOONG MEDICAL CO.,LTD. NITI-S ESOPHAGEAL COVERED STENT; ESOPHAGEAL STENT Back to Search Results
Model Number EST1808F
Device Problems Migration or Expulsion of Device (1395); Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2020
Event Type  malfunction  
Manufacturer Narrative
It was reported that approximately 18 months, two stents were found being migrated to stomach side for about 4cm.It was confirmed from the device history record that device had been manufactured with no significant issue and passed all the inspections successfully.Migration can be occurred by other company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.Based on the description, "approximately 18 months, two stents were found being migrated to stomach side for about 4cm," it is assumed that the migration has occurred due to the patient's lesion status, peristalsis of organs, and drug use.However, it is hard to exactly investigate, and analysis for this complaint since it is difficult to reconstruct the situation at the time of procedure with provided information.Through the user manual by (b)(4), it is stated that, "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: migration." this complaint is assumed that it was malfunction for device due to pressure of patient's lesion, peristalsis of organs, drug use etc complexly, and there will be continued to monitor the same or similar customer complaints.
 
Event Description
On (b)(6) 2019, an over-growth was found on the proximal side of ec1810ar above, so est1808f was additionally placed there.On (b)(6) 2019, two stents (first inserted ec1810ar stent and additionally inserted est1808f stent) were recently found being migrated to stomach side for about 4cm when checking obstruction under a scope.They were not removed and another stent (ec1812bh) was additionally placed.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NITI-S ESOPHAGEAL COVERED STENT
Type of Device
ESOPHAGEAL STENT
Manufacturer (Section D)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS  10022
Manufacturer (Section G)
TAEWOONG MEDICAL CO.,LTD.
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
KS   10022
Manufacturer Contact
lee
14, gojeong-ro
wolgot-myeon
gimpo-si, gyeonggi-do 10022
MDR Report Key10695392
MDR Text Key211873519
Report Number3003902943-2020-00084
Device Sequence Number1
Product Code ESW
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K123205
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/27/2021
Device Model NumberEST1808F
Device Catalogue NumberEST1808F
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/21/2020
Initial Date FDA Received10/19/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/28/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-