It was reported that approximately 18 months, two stents were found being migrated to stomach side for about 4cm.It was confirmed from the device history record that device had been manufactured with no significant issue and passed all the inspections successfully.Migration can be occurred by other company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.Based on the description, "approximately 18 months, two stents were found being migrated to stomach side for about 4cm," it is assumed that the migration has occurred due to the patient's lesion status, peristalsis of organs, and drug use.However, it is hard to exactly investigate, and analysis for this complaint since it is difficult to reconstruct the situation at the time of procedure with provided information.Through the user manual by (b)(4), it is stated that, "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: migration." this complaint is assumed that it was malfunction for device due to pressure of patient's lesion, peristalsis of organs, drug use etc complexly, and there will be continued to monitor the same or similar customer complaints.
|
On (b)(6) 2019, an over-growth was found on the proximal side of ec1810ar above, so est1808f was additionally placed there.On (b)(6) 2019, two stents (first inserted ec1810ar stent and additionally inserted est1808f stent) were recently found being migrated to stomach side for about 4cm when checking obstruction under a scope.They were not removed and another stent (ec1812bh) was additionally placed.There were no patient complications as a result of this event.
|