Catalog Number 11447475187 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/25/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event occurred in (b)(6).Occupation is lay user/patient.The reporter confirmed qc was acceptable.After meter testing, the reporter did not perform a plausibility check.Per product labeling, "plausibility check after glucose measurement: remove the test strip and turn it with the bottom side upwards.Compare the reaction area on the bottom side of the strip with the color field scale on the label of the test strip container.The color of the reaction area must approximately match the color assigned to your measurement result.If there is a great deviation, perform a meter function check." the reporter's test strips and meter were requested for return.The product has not been received at this time and is not expected to be returned.If the product is returned in the future, a follow up report will be submitted.
|
|
Event Description
|
The initial reporter complained of discrepant glucose results with an accutrend plus meter serial number (b)(4) compared to two accu-chek aviva meters.The reporter performed side by side meter comparison testing with the accutrend plus meter and two accu-chek aviva meters.During comparison testing, the reporters "blood was applied on one strip first, then on the other strip straight after." the reporter confirmed the test strips used for comparison testing were visually ok.The reporter's glucose result with the accutrend plus meter was 2.8 mmol/l.The reporter's glucose result with an accu-chek aviva was 5.2 mmol/l.The reporter's glucose result with another accu-chek aviva was "5.2 or 5.3 mmol/l.".
|
|
Manufacturer Narrative
|
Customer samples and retention samples glucose lot 37996005 were measured with controls gkl 60 lot 346092-99 /gkl 180 lot 346093-99 and 2 edta blood on customer device and reference devices compared to laboratory method.Results: the mean of the measurements with returned samples glucose lot 37996005 test strips and 1st edta blood sample is as follows: - customer instrument: 82.9mg/dl - retention instrument: 88 mg/dl the mean of the measurements with returned samples glucose lot 37996005 test strips and 2nd edta blood sample is as follows: - customer instrument: 140.5 mg/dl - retention instrument: 136 mg/dl the mean of the measurements with retention samples glucose lot 37996005 test strips and 1st edta blood sample is as follows: - customer instrument: 85.9 mg/dl - retention instrument: 84.7 mg/dl the mean of the measurements with retention samples glucose lot 37996005 test strips and 2nd edta blood sample is as follows: - customer instrument: 136.3 mg/dl - retention instrument: 135.7 mg/dl the mean of the measurements with control solution gkl60 with retention test strips is as follows: - customer instrument: 4.2 mmol/l - retention instrument: 78.7 mg/dl the mean of the measurements with control solution gkl180 with retention test strips is as follows: - customer instrument: 10.7 mmmol/l - retention instrument: 196 mg/d ranges glu control solutions: range gkl60: 3.2 - 5.4 mmol/l 58-98 mg/dl range gkl180: 9.5 - 12.9 mmol/l 172 ¿ 232 mmol/l the mean of the measurements with laboratory method (cobas) is as follows: - blood 1: 81.7 mg/dl - blood 2: 133.4 mg/dl all measurements with retention test strips on the reference instrument and the customer instrument are within the control ranges.No product problem was found.Medwatch fields d9 and h3 were updated.
|
|
Search Alerts/Recalls
|