The initial reporter received questionable elecsys tsh assay results for one patient tested on a cobas 8000 e 801 module serial number (b)(4).The patient¿s initial tsh result was reported outside the laboratory, and the physician questioned the result.The customer provided the patient¿s sample for investigation, and the patient¿s sample was tested on an abbott architect and a cobas 8000 e 801 module.
|
The customer's calibration and qc data were requested but not provided.The patient's sample was provided for an investigation.The investigation reproduced the customer's tsh result.Upon further investigation of the sample, an interfering factor against a component of the reagent was not detected.Further clarification of the observed discrepancy for the tsh assay is not possible with available methods and the current state of the art.Product labeling states, "for diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings." the investigation did not identify a product problem.The cause of the event could not be determined.
|