Investigation completed on a smiths medical syringe infusion pumps|medfusion 3500 pumps.The complaint of motor rate error was revealed in the event history, along with validated during flow testing.This was isolated to the cause of the main board knocked out of the grove on the extrusion.It was believed to be customer induced.Action was taken to install a main board into the extrusion.Summary of all maintenance included : pump is set to default configuration id 1a12 to perform tests.Replaced damaged bottom case and right plunger case.Replaced battery due to high cpi.Cleaned, greased and calibrated the device, performed power up process, occlusion test and all functional tests which passed.Returned l-bracket with the pump.
|