The initial reporter stated they received false negative results for one patient sample tested with the elecsys rubella igg immunoassay on a cobas 6000 e 601 module.No incorrect results were reported outside of the laboratory.The sample was initially tested using the elecsys rubella igg immunoassay, resulting with a value of 8.16 iu/ml (non-reactive).The sample was repeated on 28-sep-2020 using the elecsys rubella igg immunoassay, resulting with a value of 7.63 iu/ml (non-reactive).The sample was tested using a rubella igg elisa assay manufactured by (b)(4), resulting with a value of 69 iu/ml (positive).The sample was also tested with an unknown chemiluminescence method and the rubella igg result is always positive.The serial number of the e 601 analyzer is (b)(4).
|