• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number TMICL13.2
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cataract (1766); Capsular Bag Tear (2639)
Event Date 09/15/2020
Event Type  Injury  
Manufacturer Narrative
The lens was returned in liquid in lens vial.Visual inspection found the haptics torn.(b)(4).No similar complaint was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated the surgeon inserted and removed a 13.2mm tmicl13.2 implantable collamer lens, -13.00/+3.5/094 (sphere/cylinder/axis), during the same surgery, due to an anterior subcapsular opacity and a capsular tear had occurred.Cataract surgery was performed and an iol lens was implanted in the ciliary sulcus.The reporter indicated the cause of the event was due to 2 pits were noted in the lens capsule (anterior) directly below the iridotomies.When the lens was inserted a capsular tear extended from the pits.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key10706991
MDR Text Key212310380
Report Number2023826-2020-02407
Device Sequence Number1
Product Code QCB
UDI-Device Identifier00841542109482
UDI-Public00841542109482
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2021
Device Model NumberTMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/24/2020
Initial Date FDA Received10/20/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age25 YR
-
-