• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE VIDEO BATON 2.0 LARGE; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE VIDEO BATON 2.0 LARGE; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0570-0382
Device Problems Display or Visual Feedback Problem (1184); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2020
Event Type  malfunction  
Manufacturer Narrative
Verathon's territory manager visited the customer and reviewed the device.He noted a pixilated image that was almost blurry but no frozen image nor loss of image.The device return is anticipated, however; at the time of the report the devices have not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event Description
The customer reported that during a patient procedure, using a glidescope video baton 2.0 large, the image was pixelated and would freeze intermittently.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
 
Manufacturer Narrative
D9, g3, g6, h2, h3, h6, h10 the glidescope video baton 2.0 large was returned to verathon for evaluation.A verathon technical service representative evaluated the returned video baton 2.0 large and confirmed the intermittent image freeze issue.The technical service representative noted obvious wear and tear to the hdmi connector which contributing to the failure.The cycle count was noted as 865.Glidescope video baton 2.0 large was scrapped and a replacement was sent to the customer.No corrective action is required at this time.Verathon will continue to monitor for trends.The glidescope operations and maintenance manual (omm) notes that "before every use, ensure that the instrument is operating correctly and has no sign of damage.Do not use this product if the device appears damaged.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE VIDEO BATON 2.0 LARGE
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
MDR Report Key10707900
MDR Text Key213700276
Report Number9615393-2020-00218
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0382
Device Catalogue Number0570-0398
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/19/2020
Initial Date Manufacturer Received 09/25/2020
Initial Date FDA Received10/20/2020
Supplement Dates Manufacturer Received11/24/2020
Supplement Dates FDA Received12/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-