Model Number 0570-0382 |
Device Problems
Display or Visual Feedback Problem (1184); Application Program Freezes, Becomes Nonfunctional (4031)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Verathon's territory manager visited the customer and reviewed the device.He noted a pixilated image that was almost blurry but no frozen image nor loss of image.The device return is anticipated, however; at the time of the report the devices have not been received by verathon.Verathon continues to investigate the reported event and a supplemental report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
|
|
Event Description
|
The customer reported that during a patient procedure, using a glidescope video baton 2.0 large, the image was pixelated and would freeze intermittently.No delay in the procedure, use of a backup device, or harm to the patient or user was reported.
|
|
Manufacturer Narrative
|
D9, g3, g6, h2, h3, h6, h10 the glidescope video baton 2.0 large was returned to verathon for evaluation.A verathon technical service representative evaluated the returned video baton 2.0 large and confirmed the intermittent image freeze issue.The technical service representative noted obvious wear and tear to the hdmi connector which contributing to the failure.The cycle count was noted as 865.Glidescope video baton 2.0 large was scrapped and a replacement was sent to the customer.No corrective action is required at this time.Verathon will continue to monitor for trends.The glidescope operations and maintenance manual (omm) notes that "before every use, ensure that the instrument is operating correctly and has no sign of damage.Do not use this product if the device appears damaged.".
|
|
Search Alerts/Recalls
|
|
|