ZIMMER BIOMET, INC. VANGUARD TI FEMORAL PS 65MM LT; PROSTHESIS KNEE Back to Search Results
Model Number N/A
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Noise, Audible (3273); Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Patient Problems Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Muscular Rigidity (1968); Pain (1994); Loss of Range of Motion (2032); Swelling (2091); Synovitis (2094); Osteolysis (2377); Reaction (2414); Swelling/ Edema (4577)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Medical products: unknown tibial component catalog # unknown lot # unknown.Customer has indicated that the product will not be returned because it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filled for this event: 0001825034-2020-03854.Remains implanted.
 
Event Description
It was reported that the patient underwent a knee arthroplasty.Subsequently, the patient will be considered for revision due to allergies.Patient alleges to be allergic to zirconium and chromium.No revision procedure has been reported to date.Attempt for further information has been made, but no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being filled to relay additional information, which was unknown at the time of the initial medwatch.D11: biomet ilok pri tib tray catalog # 141211 lot # 983340.Biomet finned pri stem 40mm catalog # 141314 lot # 530850.Vngd ps tib brg 11x63/67 catalog # 183621 lot # 649610.Series a asymmetric pat 31x8 catalog # 184792 lot # 711690.Simplex cement catalog # unknown lot # unknown.
 
Event Description
From additional information received, it was reported that the patient has indicated for revision due to metal allergies, pain, swelling, stiffness and synovitis.
 
Manufacturer Narrative
This follow-up report is being filled to relay additional information, which was unknown at the time of the initial medwatch.
 
Event Description
From additional information received, it was reported that the patient has indicated for revision due to metal allergies, pain, swelling, stiffness and synovitis.Also due to osteolysis, noise / clunking, loosening and migration.Further, office notes indicate that the leg is swollen at the end of the day causing it to be stiff.A baker's cyst pops out behind the knee causing sharp pain to the inside of the kneecap.
 
Manufacturer Narrative
Reported event was confirmed by review of medical records provided.Device was not returned.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.Medical records review indicates that metal allergy to mercury, nickel, tin, chromium, vanadium and zirconium.Popping and clunk with prosthesis.Femoral component is slightly rotated into flexion 5-10 degrees.Lytic lines along cement mantle, especially along femoral component.Mechanical loosening of the knee.X-rays show no evidence of hardware complication with mild join effusion and soft tissue swelling.Bone scan shows abnormal asymmetric uptake of radiopharmaceutical around the total knee hardware in the left knee.Most prominent around the femoral component with activity around the tibial component.By end of the day left leg is swollen causing the knee to be stiff.A bakers cyst pops out behind the knee by the end of the day causing sharp pain inside of the kneecap.Synovitis of the joint of the left knee likely due to reaction to the metals.No evidence of loosening and cultures are all negative.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANGUARD TI FEMORAL PS 65MM LT
Type of Device
PROSTHESIS KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10708841
MDR Text Key212267392
Report Number0001825034-2020-03853
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
PMA/PMN Number
K080204
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup,Followup,Followup
Report Date 02/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberCP113639
Device Lot Number390450
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 09/25/2020
Initial Date FDA Received10/20/2020
Supplement Dates Manufacturer Received10/20/2020
11/06/2020
02/16/2021
Supplement Dates FDA Received11/04/2020
11/18/2020
02/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight70
-
-