It was reported that during a cori tka procedure, when beginning to mill the distal femur, the bur became loose and disconnected from the drill and the robotic drill attachment (despite there being no physical complications during drill setup or error messages notifying that the bur was assembled incorrectly during the setup/calibration stage).It is believed that the bur became disconnected during milling.In order to recover, the robotic drill connection was re-established, setup screen was prompted to ensure proper insertion of the bur, followed by calibration before being able to proceed on to the femur bone removal stage.The delay was less than 30 minutes.Even though there was additional bone removal, no other complications were reported.
|
H3, h6: the real intelligence robotic drill, part number rob10013, serial number (b)(6) used in treatment was returned for evaluation.There were no visual observations identified.A functional evaluation found that the drill¿s collet parts were assembled and functioned as expected.The manual bur insertion as well as the software bur insertion function was confirmed to operate as intended.The bur was held with enough force, preventing it from falling out.There were signs the handpiece was spun without the bur inserted (bits of o-ring scattered in the collet).It is likely that the bur was not entirely secured upon bone removal, thus causing the overcut.Although the reported problem was not confirmed through a visual or functional evaluation, a factors that may have contributed to the reported symptom may have been the following: it is possible for the drill to pass calibration with an improperly inserted bur, which is a user error.The bur is secured when the user inserts the bur until a click is heard, as instructed in the bur set up screen.Refer to the real intelligence cori for knee arthroplasty for setting up, unlocking, and loading the bur.This situation is captured in the risk assessment released at the time of the complaint.The failure mode and associated risk have been anticipated within the risk file and that the documented risk level is still adequate.A review of manufacturing records indicate the software met all specifications upon release into distribution.A complaint history review found similar reports.Escalation actions applicable to the scope of the reported complaint have been identified, and it was determined that no further action is required at this time.This issue will be continuously monitored through complaint investigation and post market surveillance.Based on the investigation, no further containment or corrective actions are recommended at this time.
|