• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEGANIA MEDICAL DEVICES PVT. LTD MON-A-THERM FOLEY CATHETER WITH TEMPERATURE SENSORMON-A-THERM FOLEY CATHETER; TEMPERATURE SENSING 100% SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEGANIA MEDICAL DEVICES PVT. LTD MON-A-THERM FOLEY CATHETER WITH TEMPERATURE SENSORMON-A-THERM FOLEY CATHETER; TEMPERATURE SENSING 100% SILICONE FOLEY CATHETER Back to Search Results
Model Number 14 FRENCH
Device Problem Low Readings (2460)
Patient Problems Edema (1820); Sepsis (2067); Burn, Thermal (2530); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 10/05/2018
Event Type  malfunction  
Manufacturer Narrative
This medical device report is filed retrospectively following fda observation number 4 received by degania medical devices pvt.Ltd.During the fda inspection of 4-7 of november 2019.The observation was related to the fact that degania medical devices pvt.Ltd.Did not establish procedures for reporting mdrs to fda as a manufacturer.Till then all complaints related to the devices produced by dmd were assessed for mdr reportability and submitted as necessary to fda by degania silicone ltd.Another q medical devices division closely affiliated with dmd.Dmd capa number (b)(4) was issued to address the observation; one of the capa actions requires dmd to perform retrospective review of all the complaints received during 2018 and 2019 and submit to fda retrospective mdrs for the reportable events (with reference to original reported to fda report # 8030107-2019-00020 filed by degania silicone ltd.) in summary, until the catheter is checked, it is not possible to conclude that the catheter had any malfunction.We close this complaint as not justified.
 
Event Description
This is retrospective submission, following reassessment of our customer complaints during period 2018-2019.Customer's text: according to the reporter, it was reported that a (b)(6) yo, male patient, weighing (b)(6)kg admitted for burn over 40% of the body, currently unstable condition, in the icu being treated for burns, respiratory assistance, and sepsis.Urinary catheter inserted on (b)(6) 2018, and (b)(6) 2018, and patient became very unstable, temperature monitoring from foley measured 37.6 degree centigrade while esophageal measurement showed 41.1 degree centigrade.Patient was called down via multiple methods.Due to edema of genitals it is not possible to remove catheter at this time, but customer reported once they are able to remove it, product will be sent for failure investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MON-A-THERM FOLEY CATHETER WITH TEMPERATURE SENSORMON-A-THERM FOLEY CATHETER
Type of Device
TEMPERATURE SENSING 100% SILICONE FOLEY CATHETER
Manufacturer (Section D)
DEGANIA MEDICAL DEVICES PVT. LTD
plot no. 251, sector-6,
imt manesar
gurgaon, india 12205 0
IN  122050
Manufacturer (Section G)
DEGANIA MEDICAL DEVICES PVT. LTD.
plot no. 251, sector-6
imt manesar
gurgaon, haryana 12205 0
IN   122050
Manufacturer Contact
ratnamma nair
plot no. 251, sector-6
imt manesar
gurgaon, haryana 12205-0
IN   122050
MDR Report Key10709508
MDR Text Key214806217
Report Number3005483737-2020-00022
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K063442
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/04/2023
Device Model Number14 FRENCH
Device Catalogue Number102201101480TY
Device Lot NumberS18005350
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/18/2018
Initial Date FDA Received10/21/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-