It was reported that the stent was found being migrated.It is hard to confirm the manufacturing history of the product because the serial number was not informed to us.Migration can be occurred by other company's device as well as ours.It is affected by patient's lesion status, peristalsis of organs, and drug use in general.Based on the description, "the stent was found being migrated and it was assumed to be excreted with the feces.The stenosis might be reduced by chemotherapy leading the migration.It is assumed that the stent migration occurred due to reduction of obstruction by chemotherapy and excretion with the feces complexly, not stent malfunction.Through the user manual by (b)(4), it is stated that, "potential complications associated with the use of niti-s & comvi stent may include, but are not limited to: migration." the suspected device is not registered in the u.S., and there will be continued to monitor the same, or similar customer complaints.
|
On (b)(6) 2020, the stent was placed.Unknown: the stent was found being migrated, and it was assumed to be excreted with the feces.The stenosis might be reduced by chemotherapy leading the migration.On (b)(6) 2020, another stent was used because a stenosis was found there.There were no patient complications as a result of this event.
|