MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 2.5X53MM LT BOND CT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 2.5X53MM LT BOND CT Back to Search Results
Catalog Number 114802
Device Problems Loose or Intermittent Connection (1371); Appropriate Term/Code Not Available (3191)
Patient Problem Pain (1994)
Event Date 09/23/2020
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery - patient presented to surgeon with pain in left elbow.Upon further review and x-rays, surgeon thought the ulnar component in total elbow was loose and decided to take patient to surgery.Once left elbow was opened, the ulnar implant was indeed loose and removed.The 2 implants removed and replaced in the revision surgery are inventoried.
 
Manufacturer Narrative
Corrected data: see d.1., d.2., d.4.& d.10.Manufacturer narrative: the reason for this revision surgery was reported as loosening.The previous surgery and the surgery detailed in this event occurred 2.5 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to pain from loose component.There were no findings during the evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implants safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Manufacturer Narrative
Corrected data: manufacturer narrative: the reason for this revision surgery was reported as loosening.The previous surgery and the surgery detailed in this event occurred 2.7 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during the evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implants safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Second revision surgery - patient presented to surgeon with pain in her left elbow.Upon further review and x-rays, surgeon thought the ulnar component in her total elbow was loose and decided to take her to surgery.Once her left elbow was opened the ulnar implant was indeed loose and removed.The 2 implants removed and replaced in the revision surgery are inventoried above.
 
Manufacturer Narrative
Corrected data: b5: event described as second revision surgery in error.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event Description
Revision surgery: patient presented to surgeon with pain in her left elbow.Upon further review and x-rays, surgeon thought the ulnar component in her total elbow was loose and decided to take her to surgery.Once her left elbow was opened the ulnar implant was indeed loose and removed.The 2 implants removed and replaced in the revision surgery are inventoried above.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA 2.5X53MM LT BOND CT
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
MDR Report Key10712787
MDR Text Key212331434
Report Number1644408-2020-00942
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
PMA/PMN Number
K090473
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/13/2021
Device Catalogue Number114802
Device Lot Number082910
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/23/2020
Initial Date FDA Received10/21/2020
Supplement Dates Manufacturer Received11/03/2020
11/03/2020
04/07/2021
Supplement Dates FDA Received12/01/2020
03/05/2021
05/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
114700 LOT 804560; 114802 LOT 082910; 114802 LOT 082910
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient Weight80
-
-