Investigation conclusion: the device in question was sent to the manufacturing site in germany on 17th november 2020 for investigation upon evaluation, it could be confirmed that the device was not in original shape, and the distal tip was missing.The shaft was found to be bent, deflection of approx.3 mm.Also, there were signs of corrosion on the broken area as well as on other parts of device.The device was manufactured on 07.2007.The root cause most probably is too high force application which results in breakage on the tip.Warning of not overload the device by exerting too much force was already included in ifu.
|