G4:13oct2020.B4:15jan2021.The fse (field service engineer) evaluated the device and could not confirm the reported problem.No parts were replaced; however, during evaluation, the fse found that the device started on its own.They replaced the ui (user interface board) and the problem was resolved.The ventilator was calibrated successfully and passed all required testing.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|