MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 3-0 NDLS; PROSTHESIS, SPORTS MED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JGRKNT 1.0MM MINI 3-0 NDLS; PROSTHESIS, SPORTS MED Back to Search Results
Catalog Number 912082
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/28/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Foreign - event occurred in (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported approximately three (3) weeks ago during surgery, the plastic guide of juggerknotmini was stuck when operator try to push the anchor into the bone.The issue caused the anchor to not be deployed correctly into the bone.Surgery was completed with another device.Attempts have been made and no further information has been provided.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Visual examination of the returned product identified the juggerknot mini was returned because the clear sleeve was stuck when the operator tried to push the anchor into the bone.The clear sleeve is fully deployed to the handle.The anchor was not returned.A function check of the juggerknot mini was done and is conforming.Reported event is not related to a combination of products; therefore a compatibility review is not applicable.Medical records were not provided.Dhr was reviewed and no discrepancies relevant to the reported event were found.No device problem was found.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 1.0MM MINI 3-0 NDLS
Type of Device
PROSTHESIS, SPORTS MED
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10717398
MDR Text Key212493367
Report Number0001825034-2020-03863
Device Sequence Number1
Product Code DZL
UDI-Device Identifier00880304523418
UDI-Public(01)00880304523418
Combination Product (y/n)N
PMA/PMN Number
K140908
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number912082
Device Lot Number246690
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/05/2020
Initial Date FDA Received10/22/2020
Supplement Dates Manufacturer Received12/31/2020
Supplement Dates FDA Received12/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNI
Patient Sequence Number1
-
-