Model Number UHI-4 |
Device Problem
No Flow (2991)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not returned to olympus medical systems corp.(omsc) for evaluation.There were no further details provided.If significant additional information is received, this report will be supplemented.
|
|
Event Description
|
Olympus medical systems corp.(omsc) was informed that in unspecified timing the device gave an alarm and failed to supply co2 to the patient.Other detailed information was not provided.There was no report of patient injury associated with the event.
|
|
Manufacturer Narrative
|
This supplemental report is being submitted to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, but was returned to olympus service operation repair center (sorc).Sorc inspected the subject device and found failure in the main board and the manifold.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported phenomenon could not be conclusively determined.However, there was the possibility that the reported phenomenon was attributed to the accidental failure of the main board and/or the manifold.
|
|