ARGON MEDICAL DEVICES FIRST PICC S/L 26GA (1.9F) 0.65MM X 50CM Back to Search Results
Model Number 384232
Device Problem Positioning Problem (3009)
Patient Problem Insufficient Information (4580)
Event Date 10/05/2020
Event Type  malfunction  
Manufacturer Narrative
The device is unavailable for evaluation.However, images were provided.Based on the review of the images, the space between the luer/hub and the actual picc could not be confirmed.If the picc line was not fully attached to the hub, the leak would have been found while flushing the device prior to use.Without the actual device used available for review, the root cause cannot be determined.
 
Event Description
The picc was inserted by nicu on (b)(6) 2020.Pt is now on a peds m/s unit.Nurse called to inform me that there is a space between the luer /hub, and the actual picc.Picc is indwelling in patient.Per nurse, discussion of alternative central access was discussed in rounds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIRST PICC S/L 26GA (1.9F) 0.65MM X 50CM
Type of Device
FIRST PICC
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek rd
athens TX
Manufacturer Contact
gina liu
1445 flat creek rd, athens, tx 75751
MDR Report Key10727717
MDR Text Key215093714
Report Number1625425-2020-00696
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00886333209750
UDI-Public00886333209750
Combination Product (y/n)N
PMA/PMN Number
K972262
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number384232
Device Catalogue Number384232
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received10/23/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-